It is recommended by the International Society for Clinical Densitometry (ISCD) that every clinic performs a precision assessment of bone densitometry to evaluate the reproducibility. Bone densitometry is used for diagnosis of osteoporosis, to monitor response to treatment and to assess patients’ risk of fractures. Dual energy x-ray absorptiometry (DXA) was used to perform double scans of patients already booked for examination at the section of clinical physiology, Skåne University Hospital (SUS), Lund. 105 patients were included in the study. Measurements were made at whole body, total hip, femoral neck and lumbar spine. The whole body scans were included in an interindividual study, whereas the others were included in intraindividual studies. The reproducibility was expressed as least significant change (LSC), root mean square standard deviation (RMS SD) and coefficient of variation (%CV). The reproducibility in the study was good, with low values for LSC and RMS SD. %CV fell below the maximal values recommended by the ISCD, which are 1,8 % for total hip, 2,5 % for femoral neck and 1,9 % for lumbar spine.